Registration certificate number: jiangxi machinery note 20162540325
Products overview:
This equipment is based on noninvasive continuous positive airway pressure (NCPAP), designed specifically for pediatric respiratory support medical equipment, in pediatrics for premature infants, newborns, infants and children under the age of 14, the implementation of nasal mask ventilation (NCPAP) when oxygen concentration monitoring, regulation and function of wetting.
Alarm function:
Air input fault, electricity shortages, built-in battery low, high airway pressure, airway pressure, oxygen concentration high and low alarm, air - oxygen input unbalanced pressure alarm;
Alarm mute: no more than 120 s;
The maximum limit pressure: 60 cmH2O or less;
Performance characteristics:
Stable oxygen concentration adjustable gas supply, constant flow design;
Oxygen concentration can accurately adjust continuously;
CPAP pressure can be stable, precise adjustment;
Inhaled air and humid, and the gas constant temperature automatic control;
The real-time monitoring gas oxygen density;
A variety of alarm functions, to ensure safe and reliable ventilation;
A built-in battery, ensure continuous power supply.
Technical parameters:
Inhaled oxygen concentration regulation and monitoring scope: 21% to 100%.
Flow control and monitoring scope: 2 L/min - 20 L/min;
Airway pressure regulating range: 2 cmH2O - 20 cmH2O
CmH2O airway pressure monitoring scope: 0-60 cmH2O